Dionex AERS 500 Carbonate Manual

The Dionex AERS 500 Carbonate is an advanced electrolytic reagent suppressor for ion chromatography systems, designed for carbonate/bicarbonate eluents. It features automatic regeneration, low noise operation, high suppression capacity, and compatibility with various Dionex IC systems. Below are key sections for safety, features, installation, operation, maintenance, and troubleshooting.

1. Features at a Glance 2. Installation 3. First-Time Setup 4. Specifications 5. Controls and Settings 6. Connecting to IC System 7. Operation Procedures 8. Maintenance Guidelines 9. Care & Cleaning 10. Regulatory Information 11. Troubleshooting

Features at a Glance

Key components: Suppressor module, Regeneration cartridge, Power supply unit, Control interface, Fluid connections, Status indicators.

FeatureDescription
Suppression CapacityHigh capacity for carbonate/bicarbonate eluents
Regeneration MethodAutomatic electrolytic regeneration
CompatibilityWorks with Dionex ICS-2000, ICS-2100, ICS-3000 systems
Noise LevelLow electrical noise for improved detection
Flow Rate Range0.1 to 2.0 mL/min
Pressure LimitMaximum 200 psi (13.8 bar)
Power Requirements24V DC power supply
Status IndicatorsLED indicators for power, operation, error
InterfaceRS-232 communication port
DimensionsStandard 19-inch rack mountable

Installation

Unpack and inspect contents. Ensure proper laboratory environment.

  1. Mount unit: Install in 19-inch rack or stable bench surface.
  2. Connect power: Use provided 24V DC power supply.
  3. Connect to IC system: Attach eluent inlet and outlet tubing.
  4. Connect waste: Attach waste tubing to regeneration waste outlet.
  5. Connect communication: Attach RS-232 cable if computer control required.

WARNING! Ensure proper grounding. Use only specified power supply to prevent damage.

First-Time Setup

Power on unit; follow initialization procedure.

  1. Prime system: Flush with deionized water for 10 minutes.
  2. Condition suppressor: Run with carbonate eluent for 30 minutes.
  3. Verify operation: Check pressure and baseline stability.
  4. Configure settings: Set flow rate and suppression current via control panel.
  5. Connect to IC system: Ensure proper integration with chromatography software.

CAUTION! Always use high-purity water and reagents to prevent contamination.

Specifications

Controls and Settings

Access controls via front panel or software interface.

Front Panel: POWER switch, STATUS indicators, FLOW RATE adjustment, SUPPRESSION CURRENT control.

Software Control: Current setting, Flow monitoring, Pressure monitoring, Error logging, Calibration parameters.

Operation Modes: Continuous operation, Standby mode, Purge mode, Calibration mode.

Connecting to IC System

Compatible with Dionex ICS series chromatography systems.

  1. Turn off all equipment before connecting.
  2. Connect eluent inlet: From IC pump to suppressor inlet.
  3. Connect eluent outlet: From suppressor outlet to detector.
  4. Connect regeneration lines: Water inlet and waste outlet.
  5. Connect power and communication cables.
  6. Verify all connections are secure and leak-free.

Tip: Use Dionex PEEK tubing and fittings for optimal performance.

Operation Procedures

Standard operating protocol for carbonate analysis.

  1. Startup: Power on unit, allow 5-minute initialization.
  2. Set parameters: Adjust flow rate and suppression current as required.
  3. Begin analysis: Start IC system and data collection.
  4. Monitor performance: Watch pressure and baseline stability.
  5. Shutdown: Flush with deionized water for 15 minutes before turning off.
  6. Storage: If not used for extended periods, store with water-filled channels.

WARNING! Never run suppressor dry. Always maintain fluid flow during operation.

Maintenance Guidelines

Regular maintenance ensures optimal performance and longevity.

Maintenance Log: Keep record of all maintenance activities and performance data.

Care & Cleaning

Proper cleaning procedures for maintenance and contamination removal.

  1. Routine cleaning: Flush with 0.1M HCl for 30 minutes, then deionized water.
  2. Deep cleaning: Use manufacturer-recommended cleaning solutions.
  3. Storage cleaning: Flush thoroughly with deionized water before storage.
  4. Exterior cleaning: Use mild detergent and soft cloth.

CAUTION! Use only approved cleaning solutions. Never use organic solvents.

Regulatory Information

CE Mark: Complies with European Union safety standards. UL Listed: Meets UL 61010-1 safety requirements. RoHS Compliant: Meets Restriction of Hazardous Substances directive. Copyright © 2024 Thermo Fisher Scientific. For technical support, contact Thermo Fisher Scientific at 1-800-532-4752 or visit www.thermofisher.com.

Troubleshooting

SymptomPossible CauseCorrective Action
High backpressureClogged suppressorClean with recommended procedure; check for particulates
Noisy baselineContaminationFlush system; replace eluent; check water purity
Low suppressionExhausted cartridgeReplace regeneration cartridge; verify current settings
No powerPower supply issueCheck power connections; verify power supply output
Communication errorCable/interface issueCheck RS-232 connection; verify software settings
LeaksLoose fittingsTighten all connections; replace damaged tubing

Reset Procedure: Power cycle unit; reinitialize settings.

Technical Support: 1-800-532-4752; www.thermofisher.com

Dionex AERS 500 Carbonate Manual © - All Rights Reserved 2025