The Dionex HPLC (High Performance Liquid Chromatography) system is a sophisticated analytical instrument designed for precise separation, identification, and quantification of chemical compounds. This manual covers the UltiMate 3000 series with quaternary pump, autosampler, column compartment, and various detectors including DAD, FLD, and CAD. Below are key sections for safety, features, installation, setup, operation, maintenance, and troubleshooting.
Key components: Quaternary pump, autosampler, column compartment, detector array, Chromeleon software interface.
| Feature | Description |
|---|---|
| Quaternary Pump | Four-solvent capability with high-pressure mixing |
| Autosampler | Precise sample injection with temperature control |
| Column Compartment | Temperature-controlled column environment |
| DAD Detector | Diode Array Detector for multi-wavelength detection |
| FLD Detector | Fluorescence Detector for sensitive analysis |
| CAD Detector | Charged Aerosol Detector for universal detection |
| Chromeleon Software | Comprehensive data acquisition and analysis |
| Degasser | Online solvent degassing system |
| Column Heater | Precise temperature control up to 80°C |
| Sample Tray | Capacity for 100+ vials with barcode reading |
Unpack and inspect all components. Place on stable, level surface.
WARNING! Handle columns with care; avoid dropping or shocking. Ensure all fittings are properly tightened.
Power on system; initialize Chromeleon software.
CAUTION! Use HPLC-grade solvents only. Wear appropriate personal protective equipment.
Access system controls via Chromeleon software interface or front panel displays.
Software Controls: Method editor, sequence setup, real-time monitoring, data analysis.
Pump Settings: Flow rate, pressure limits, gradient program, solvent composition. Autosampler: Injection volume, wash cycles, needle position, temperature. Detector: Wavelength selection, bandwidth, acquisition rate, reference wavelength. Column Oven: Temperature setpoint, equilibration time, temperature ramp.
Proper connection sequence: Solvent reservoirs → Degasser → Pump → Autosampler → Column → Detector → Waste.
Tip: Use finger-tight fittings; avoid over-tightening which can damage threads.
Create and optimize HPLC methods for specific applications.
WARNING! Always equilibrate column with starting mobile phase before analysis.
Using Chromeleon software for chromatographic data processing.
Regular maintenance schedule for optimal performance.
CAUTION! Flush system with appropriate solvents before storage. Use only HPLC-grade solvents for cleaning.
CE marked: Complies with European Union safety standards. FDA 21 CFR Part 11: Software complies with electronic records requirements. RoHS compliant: Meets Restriction of Hazardous Substances directive. For technical support, contact Thermo Fisher Scientific at 1-800-532-4752 or visit www.thermofisher.com.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| High pressure | Clogged column/filter | Replace inline filter; flush or replace column. |
| Peak tailing | Column degradation | Replace column; check mobile phase pH. |
| Retention time shift | Mobile phase/temperature | Prepare fresh mobile phase; check column temperature. |
| No peaks | Detector lamp/settings | Check detector lamp life; verify wavelength settings. |
| Baseline noise | Contamination/bubbles | Degas mobile phase; clean system; replace lamp. |
| Leaks | Loose fittings/seals | Tighten fittings; replace pump seals or tubing. |
| Poor resolution | Method/column issues | Optimize gradient; replace column; adjust flow rate. |
Reset: Power cycle system; reinitialize Chromeleon software if needed.
Service: 1-800-532-4752; www.thermofisher.com